Product Description
HPHT Curing Autoclave 7360
HPHT Curing Autoclave is designed to prepare testing cement sample in accordance with API test procedures.
Feature
Curing autoclave HTD7360 has the following characteristics:
1.The controller can be multi point PID control heating and heat preservation process.
2.A variety of pressure limiting control, high pressure limit through the front panel on the pressure regulator
3.The equipment of pump pressure adjustable automatic maintenance equipment, pressure, when the pressure is lower than the limit of pressure, pump automatically start booster.
Overheat alarm
If the kettle body temperature exceeds the set temperature, the machine will automatically stop heating and begins to cool, you must press the alarm reset switch on the front panel, the heater can work again.
Product Parameters
Temperature control |
external 3300W heater, and the internal and external thermocouples are connected to the controller, the controller can be multi point PID control heating and heat preservation process. |
Pressure system |
equipment supercharged by air driven booster pump pressure relief, equipment using air switch, the pressure is controlled by pressure transmitter and controller accurate control, precision is ±0.2MPa. |
Maximum pressure |
6000 psi/41 Mpa |
Maximum temperature |
500°F/260°C |
Power of heater |
3300W |
Heater type |
external, 2 tablets and tropical |
The heater control |
PID temperature control |
Rated Frequency
Input pressure |
AC220±10% 50Hz |
Total power |
3.5 KW |
Mechanical
Height |
20 in/68 cm |
Width |
21.5 in/60 cm |
Length |
19 in/68cm |
Weight |
150 lb/68 kg |
Environmental requirements
Temperature |
32°F - 105°F/0°C - 40°C |
Humidity |
0 - 95% |
Compressed air and water
Compressed air |
0.69MPa |
Water |
0.5MPa |
Detailed Photos
Services
1. Warranty 1 year of replacement and return.
2. Providing factory visit and study, technical training.