SYD-261 Pensky-Martens Closed-Cup Flash Point Tester
Summary
This instrument is designed and manufactured as per standards GB/T 261-2008 Determination of
Flash Point - Pensky-Martens Closed Cup Method and ASTM D93 Standard Test Method for
Flash Point by Pensky-Martens Closed Cup Tester. It is used to make determination of the closed
cup flash point of the petroleum products.
I. Main technical features
1. This instrument is designed as per GB/T261 and ASTM D93.
2. The heating power is continuously adjustable. The power is shown visually by voltmeter.The
temperature control mode is advanced and reasonable.
3. The structure is designed small and exquisite.The stainless-steel table board is beautiful and
easy to clean.
4. Operation is easy.Test results are precise.It has been chosen as the instrument to test the closed
cup flash point by many metering and detecting institutes.
II. Main technical specifications
1. Power supply:AC (220±10%)V, 50Hz.
2. Heating device::The heating power is continuously adjustable from 0 W to 600W.
3. Heating rate::(1~12)ºC/min;Controllable and adjustable.
4. Stirring rate: Procedure A: (90~120)RPM, Procedure B: (250±10)RPM
5. Oil cup:(1) Inner diameter: 50.7mm~50.8mm.
(2) Depth: 55.7mm~56.0mm
(3) The scribed line depth of capacity of testing oil: 33.9mm~34.3mm.
(4) Capacity of testing oil:about 70ml
6. Igniting device:(1) Igniting source: gas(
or other civilian fuels,the same below)
(2) Electric ignition. Gas flame diameter:3.2mm~4.8mm
7. Thermometers:Mercury-in-glass thermometer.Specifications are as below:
(1)Scale -5ºC~110ºC, division 0.5ºC.
(2)Scale 20ºC~150ºC, division 1ºC.
(3)Scale 90ºC~370ºC,division 2ºC.
8. Ambient temperature:≤ 35ºC
9. Relative humidity:≤ 85%
10.Maximum power consumption:650W
Parameter Names |
Details |
Ignition method |
Autoignition |
Electric heating power |
400W |
Thermoregulation |
Thermostat using SCR |
Heating method |
Thermostat using SCR |
LPG tank |
5L |